| Intro: |
Automatic sonic nozzle gas meter test bench G6-100 is an equipment to test industrial gas meter, the flow range 0.06m3/h¡«160m3/h. It is an ideal testing equipment for gas meter factory and R&D institute.
MAIN TECHNICAL 1. Test bench accuracy: ¡Ü0.5% 2. Sonic Nozzle accuracy: ¡Ü0.2% 3. Range of pressure sensor: (0¡«100) kPa, accuracy: ¡Ü0.2% (imported from UK or Germany) 4. Range of temperature sensor: 273.15¡«323.15K, accuracy : ¡Ü0.2% (imported from Germany) 5. The test bench adopts sonic nozzle as standard meter (total 14pcs sonic nozzles inside), it can test gas meters: G6¡¢G10¡¢G16¡¢G25¡¢G40¡¢G65¡¢G100. Testing points (Qmax, 0.2Qmax, Qmin): 0.06m3/h, 0. 1 m3/h, 0.16 m3/h, 0.25 m3/h, 0.40 m3/h, 0.65 m3/h, 1 m3/h, 2 m3/h, 3.2 m3/h, 5 m3/h, 8 m3/h, 10 m3/h, 13 m3/h, 16 m3/h, 20 m3/h, 25 m3/h, 32 m3/h, 40 m3/h, 65 m3/h, 100 m3/h, 160 m3/h, total 21 testing points. 6. Flow accuracy: £¼0.1% 7. Crystal accuracy: £¼0.01%/daily 8. The test bench can test 1pc gas meter each time 9. Working invirement£º environment temperature: 20¡æ¡À5¡æ; relevant humidity: 60%¡«70%; atmospheric pressure: 86kPa¡«106kPa 10. Power supply: 380V, 50Hz 11. The testing process is controlled by PC, meters are clipped manually.
|